User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: Sub HP 54501A: TMA I3E DATE: 19-Jul-94 AUTHOR: User Contributed REVISION: 1.0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 2 NUMBER OF LINES: 41 CONFIGURATION: Tektronix SG5030 ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK+ B 1.002 ASK- R N P J F W 1.003 5030 * S # # 1.004 HEAD [27][91]1m{Time Measurement Accuracy} 1.005 DISP Connect the SG 5030 to the UUT channel 1 input using 1.005 DISP the amplitude leveling head and a 50ê termination.[7] 1.006 IEEE :MENU CHAN;*RST;*RST;:DISP:GRAT GRID 1.007 IEEE :ACQ:TYPE AVER;:ACQ:COUN 16 1.008 IEEE :TIM:RANG 1.8E-6;:CHAN:RANG .4 1.009 5030 1uT 282.8mVP -D S 1.010 IEEE :MENU TRIG;:TRIG:LEV 0MV;:SYST:KEY 39 1.011 DISP 5. Adjust trigger level so the positive edge of the 1.011 DISP signal crosses a graticule line exactly at center 1.011 DISP screen or some other convenient reference.[7] 1.012 IEEE :MENU TIM;:SYST:KEY 11;:TIM:DEL .5MS;:SYST:KEY 39 1.013 DISP 7. Use the knob to set the positive edge of the signal 1.013 DISP to center screen(reference).[7] 1.014 MATH MEM1 = 500 1.015 IEEE :TIM:DEL?[D1000][I] 1.016 MATH MEM = 1E6 * MEM 1.017 MEME 1.018 MEMC T 0.03U # # 2.001 IEEE :MENU CHAN;*RST;*RST;:DISP:GRAT GRID;:SYST:KEY 39 2.002 5030 * S 2.003 END